mesoSPIM
The mesoSPIM is a cutting-edge axially scanned light-sheet microscope designed specifically for imaging large cleared samples — from whole mouse brains to entire central nervous systems — with near-isotropic resolution in as little as 5–8 minutes. As an open-hardware instrument, it is compatible with virtually any clearing or expansion protocol.
A vertical rotation axis enables 360° sample rotation without changing the direction of gravity, preventing soft-tissue distortions. Dual-sided illumination combined with the axially scanned light-sheet technology minimizes shadowing artifacts and delivers a uniform 6.5 μm voxel size across the entire field of view. Kinematic mounts and magnetic sample holders make sample exchange fast and reproducible.
Compatible clearing techniques include CLARITY, X-CLARITY, CUBIC-X, BABB, iDISCO, and MASH. Image analysis is supported by a full software suite: IMARIS, Huygens, AMIRA, and Wiener deconvolution. Detailed hardware documentation is available on the mesoSPIM GitHub repository.
- Travel range: 44.5 × 44.5 × 100 mm
- Voxel size: 6.5 μm (uniform across FOV)
- Scan time: 5–8 min per sample
- Rotation: 360° vertical axis
- Lasers: 405 · 488 · 561 · 594 · 647 nm
- Camera: Kinetix sCMOS · 10.24 MP · 96% QE
- Workstation: 1 TB RAM · 2× RTX A6000 48 GB
- Software: IMARIS · Huygens · AMIRA · Wiener
- Base: mesoSPIM V5 — custom MDIBL variant (upgraded mid-2023)
- Olympus SWTLU-C wide-field tube lens (180 mm focal length)
- Horizontal detection axis
- ZWO 7 × 2″ motorized filter wheel
- 2 × PI L-509 linear stages — 100 mm travel — sample XY translation
- 1 × M-060.DG rotation stage — 360° vertical axis sample rotation
- 1 × PI M-111.2DG cuvette stage — translates cuvette wall to act as motorized correction collar for objectives
- 1 × PI M-406 focus stage — controls detection path position
- Travel envelope: 44.5 × 44.5 × 100 mm
- Controller A: PI C-884.6DC — sample XYZ, rotation, cuvette & focus stages (6 axes)
- Controller B: PI C-863.11 — objective turret stage
- ⚠ When switching objectives, cuvette stage moves automatically to safe position before turret rotates
- Dual-sided excitation via fiber-coupled multiline laser combiner
- 120 mW · 405 nm LuxX (analog mod. up to 1.5 MHz · digital up to 150 kHz)
- 150 mW · 488 nm LuxX (analog mod. up to 1.5 MHz · digital up to 150 kHz)
- 150 mW · 561 nm DPSS (analog & digital mod. up to 2.5 MHz)
- 100 mW · 594 nm DPSS (analog & digital mod. up to 2.5 MHz)
- 140 mW · 647 nm LuxX (analog mod. up to 1.5 MHz · digital up to 150 kHz)
- Clean-up filters on all LuxX lasers
- Fast fiber switch · dual fiber output · breadboard mounted
- 2 × point-source fiber couplers · 3 m SM PM fibers · FC/APC
- One open slot for additional wavelength
- Illumination objective: Nikon AF-S 50 mm f/1.4 G (ref: 91863)
- ETL lens: Optotune EL-16-40-TC-VIS-5D-1-C
- ETL driver: Optotune EL-E-4i
- ETL cables: Optotune CAB-6-300
- Galvo scanners: Thorlabs GVS011
- Mirrors: 2 × elliptic 2″ (Thorlabs BBE2-E02)
- Objective turret: PI MICOS L-611 heavy-duty rotation stage (detection path)
- ET480/30m — Chroma
- ET525/30m — Chroma
- ET540/30m — Chroma
- ET605/40m — Chroma (custom design)
- ET625/40m — Chroma
- ET665lp — Chroma
- Ref: 410830-1768-000
- Sensor: Scientific CMOS · 29.4 mm diagonal (20.8 × 20.8 mm)
- Pixel count: 3200 × 3200 = 10.24 MP
- Pixel size: 6.5 μm × 6.5 μm
- Bit depth: 8-bit (speed) · 12-bit (sensitivity) · 16-bit (dynamic range)
- Frame rate: 5.2 – 498 fps
- Quantum efficiency: 96% at 600 nm
- 40 × 40 × 40 mm³ standard cuvette
- 50 × 50 × 50 mm³ large cuvette
- Temperature-controlled immersion cuvette
- Workstation: Lenovo ThinkStation P920 · Windows 11 Pro 64-bit
- CPU: 2 × Intel Xeon Platinum 8280 · 2.70 GHz (4.00 GHz boost) · 28 cores · 38.5 MB cache
- RAM: 1 TB DDR4 2933 MHz
- GPU: 2 × NVIDIA RTX A6000 48 GB GDDR6 (NVLink)
- Acquisition: mesoSPIM-control (MDIBL fork — mdibl-development-2) — Z-stack · Multi-view · Tiling & stitching · Multi-color · Time-lapse
- Analysis: IMARIS · Huygens · AMIRA · Wiener deconvolution
These are dipping objectives designed for organic solvents. The correction collar is motorized via cuvette wall translation — the cuvette stage moves automatically when switching between objectives to maintain optimal image quality.
| Objective | Magnification | NA | Working Distance | Reference |
|---|---|---|---|---|
| MI Plan 1.1× | 1.1× | 0.1 | <17 mm | 130-133-611 |
| MI Plan 4.0× | 4.0× | 0.35 | <16 mm | 130-133-602 |
| MI Plan 12× | 12× | 0.53 | <10.9 mm | 130-133-605 |
✦ Custom Chroma design