The mesoSPIM

← All Instruments
Light Sheet Fluorescence Microscope

mesoSPIM

Light Sheet Upright Dual-sided Illumination Screening Phenotyping 3D Imaging Optical Sectioning Tiling & Stitching Cleared Samples Whole Organisms Whole Organs

The mesoSPIM is a cutting-edge axially scanned light-sheet microscope designed specifically for imaging large cleared samples — from whole mouse brains to entire central nervous systems — with near-isotropic resolution in as little as 5–8 minutes. As an open-hardware instrument, it is compatible with virtually any clearing or expansion protocol.

A vertical rotation axis enables 360° sample rotation without changing the direction of gravity, preventing soft-tissue distortions. Dual-sided illumination combined with the axially scanned light-sheet technology minimizes shadowing artifacts and delivers a uniform 6.5 μm voxel size across the entire field of view. Kinematic mounts and magnetic sample holders make sample exchange fast and reproducible.

Compatible clearing techniques include CLARITY, X-CLARITY, CUBIC-X, BABB, iDISCO, and MASH. Image analysis is supported by a full software suite: IMARIS, Huygens, AMIRA, and Wiener deconvolution. Detailed hardware documentation is available on the mesoSPIM GitHub repository.

mesoSPIM Light Sheet Microscope at MDIBL LMF
At a glance
  • Travel range: 44.5 × 44.5 × 100 mm
  • Voxel size: 6.5 μm (uniform across FOV)
  • Scan time: 5–8 min per sample
  • Rotation: 360° vertical axis
  • Lasers: 405 · 488 · 561 · 594 · 647 nm
  • Camera: Kinetix sCMOS · 10.24 MP · 96% QE
  • Workstation: 1 TB RAM · 2× RTX A6000 48 GB
  • Software: IMARIS · Huygens · AMIRA · Wiener
Hardware
System Specifications
  • Base: mesoSPIM V5 — custom MDIBL variant (upgraded mid-2023)
  • Olympus SWTLU-C wide-field tube lens (180 mm focal length)
  • Horizontal detection axis
  • ZWO 7 × 2″ motorized filter wheel
  • 2 × PI L-509 linear stages — 100 mm travel — sample XY translation
  • 1 × M-060.DG rotation stage — 360° vertical axis sample rotation
  • 1 × PI M-111.2DG cuvette stage — translates cuvette wall to act as motorized correction collar for objectives
  • 1 × PI M-406 focus stage — controls detection path position
  • Travel envelope: 44.5 × 44.5 × 100 mm
  • Controller A: PI C-884.6DC — sample XYZ, rotation, cuvette & focus stages (6 axes)
  • Controller B: PI C-863.11 — objective turret stage
  • ⚠ When switching objectives, cuvette stage moves automatically to safe position before turret rotates
  • Dual-sided excitation via fiber-coupled multiline laser combiner
  • 120 mW · 405 nm LuxX (analog mod. up to 1.5 MHz · digital up to 150 kHz)
  • 150 mW · 488 nm LuxX (analog mod. up to 1.5 MHz · digital up to 150 kHz)
  • 150 mW · 561 nm DPSS (analog & digital mod. up to 2.5 MHz)
  • 100 mW · 594 nm DPSS (analog & digital mod. up to 2.5 MHz)
  • 140 mW · 647 nm LuxX (analog mod. up to 1.5 MHz · digital up to 150 kHz)
  • Clean-up filters on all LuxX lasers
  • Fast fiber switch · dual fiber output · breadboard mounted
  • 2 × point-source fiber couplers · 3 m SM PM fibers · FC/APC
  • One open slot for additional wavelength
  • Illumination objective: Nikon AF-S 50 mm f/1.4 G (ref: 91863)
  • ETL lens: Optotune EL-16-40-TC-VIS-5D-1-C
  • ETL driver: Optotune EL-E-4i
  • ETL cables: Optotune CAB-6-300
  • Galvo scanners: Thorlabs GVS011
  • Mirrors: 2 × elliptic 2″ (Thorlabs BBE2-E02)
  • Objective turret: PI MICOS L-611 heavy-duty rotation stage (detection path)
  • ET480/30m — Chroma
  • ET525/30m — Chroma
  • ET540/30m — Chroma
  • ET605/40m — Chroma (custom design)
  • ET625/40m — Chroma
  • ET665lp — Chroma
  • Ref: 410830-1768-000
  • Sensor: Scientific CMOS · 29.4 mm diagonal (20.8 × 20.8 mm)
  • Pixel count: 3200 × 3200 = 10.24 MP
  • Pixel size: 6.5 μm × 6.5 μm
  • Bit depth: 8-bit (speed) · 12-bit (sensitivity) · 16-bit (dynamic range)
  • Frame rate: 5.2 – 498 fps
  • Quantum efficiency: 96% at 600 nm
  • 40 × 40 × 40 mm³ standard cuvette
  • 50 × 50 × 50 mm³ large cuvette
  • Temperature-controlled immersion cuvette
  • Workstation: Lenovo ThinkStation P920 · Windows 11 Pro 64-bit
  • CPU: 2 × Intel Xeon Platinum 8280 · 2.70 GHz (4.00 GHz boost) · 28 cores · 38.5 MB cache
  • RAM: 1 TB DDR4 2933 MHz
  • GPU: 2 × NVIDIA RTX A6000 48 GB GDDR6 (NVLink)
  • Acquisition: mesoSPIM-control (MDIBL fork — mdibl-development-2) — Z-stack · Multi-view · Tiling & stitching · Multi-color · Time-lapse
  • Analysis: IMARIS · Huygens · AMIRA · Wiener deconvolution
Optics
Detection Objectives — Miltenyi Biotec (LaVision BioTec)

These are dipping objectives designed for organic solvents. The correction collar is motorized via cuvette wall translation — the cuvette stage moves automatically when switching between objectives to maintain optimal image quality.

Objective Magnification NA Working Distance Reference
MI Plan 1.1× 1.1× 0.1 <17 mm 130-133-611
MI Plan 4.0× 4.0× 0.35 <16 mm 130-133-602
MI Plan 12× 12× 0.53 <10.9 mm 130-133-605
Optical Configuration
Filter Sets — Interactive Viewer
DAPI / AF350 Chroma ET480/30m Ex ~405 nm · Em ~480 nm
eGFP / AF488 Chroma ET525/30m Ex ~488 nm · Em ~525 nm
eYFP / Venus Chroma ET540/30m Ex ~488 nm · Em ~540 nm
AF568 / tdTomato Chroma ET605/40m ✦ Ex ~561 nm · Em ~605 nm
mCherry / AF594 Chroma ET625/40m Ex ~594 nm · Em ~625 nm
AF647 / Cy5 Chroma ET665lp Ex ~647 nm · Em >665 nm

✦ Custom Chroma design

Ready to image your cleared samples?
Contact Dr. Bonnet to discuss your project — on-site visits, remote imaging (ship fixed samples), and collaborative arrangements are available.
Get in touch